<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>45243</ErrorNumber>
<ErrorFamily>Product Support</ErrorFamily>
<ErrorMessage>
PRD: Head Power Module safety circuit failure
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Magnus_800</Text></ListItem>
<ListItem><Text>Magnus_VLF</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The TH3 power module contains a safety circuit that shunts energy from the
internal storage capacitors, in order to ensure that the head is actually
powered off as quickly as possible.  Each time the head is powered on, the
safety circuit is checked to make sure it is functioning properly.  If it
is determined that it is NOT functioning properly, this error will be
generated and the head will not be powered on.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><UnnumberedList>
<ListItem><Text>
Try resetting the device and see if the behaviour is repeated
</Text></ListItem>
<ListItem><Text>
If repeated, replace the TH3 power module
</Text></ListItem>
</UnnumberedList></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
Hardware problem in TH3 power module
</Cause>
<LastModified>
23-Feb-2005
</LastModified>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 1 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 1 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllParameters.htm</LinkAddress><LinkTitle>Firmware 1 Parameter Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Thu Jul 17 07:23:45 2008 PST
</Updated>
<Links>
</Links>
</Error>